| Title | Collision cross section calibrants for negative ion mode traveling wave ion mobility-mass spectrometry. |
| Publication Type | Journal Article |
| Year of Publication | 2015 |
| Authors | Forsythe, JG, Petrov, AS, Walker, CA, Allen, SJ, Pellissier, JS, Bush, MF, Hud, NV, Fernández, FM |
| Journal | Analyst |
| Volume | 140 |
| Issue | 20 |
| Pagination | 6853-61 |
| Date Published | 2015 Sep 28 |
| ISSN | 1364-5528 |
| Abstract | Unlike traditional drift-tube ion mobility-mass spectrometry, traveling-wave ion mobility-mass spectrometry typically requires calibration in order to generate collision cross section (CCS) values. Although this has received a significant amount of attention for positive-ion mode analysis, little attention has been paid for CCS calibration in negative ion mode. Here, we provide drift-tube CCS values for [M - H](-) ions of two calibrant series, polyalanine and polymalic acid, and evaluate both types of calibrants in terms of the accuracy and precision of the traveling-wave ion mobility CCS values that they produce. |
| DOI | 10.1039/c5an00946d |
| Alternate Journal | Analyst |
| PubMed ID | 26148962 |